Chapter 1 · Theory

Photons in, electrons out

The photoelectric effect and the one equation behind every XPS spectrum.

X-ray photoelectron spectroscopy (XPS, also called ESCA) shines X-ray photons of a known energy onto a sample. A photon is absorbed by an atom and gives all of its energy to one electron. If that energy is larger than what holds the electron in the atom, the electron leaves: it is a photoelectron.

+holehν (X-ray)photoelectron, KEVacuum levelFermi level2p2s1shνBEφKE
Left: an X-ray photon ejects a core electron and leaves a hole. Right: the energy balance — the photon energy hν pays the binding energy BE, the work function φ, and what is left is the kinetic energy KE.

We cannot measure directly how strongly the electron was bound. What the instrument measures is the kinetic energy (KE) of the electron as it arrives in the analyser. Energy is conserved, so:

KE = hν − BE − φ

  • hν — the energy of the X-ray photon (1486.6 eV for Al Kα).
  • BE — the binding energy: the energy needed to remove the electron from its level. This is what we want.
  • φ — the work function of the spectrometer, a small constant (≈ 4–5 eV) found by calibration.
  • KE — the kinetic energy measured by the analyser.

Rearranged, BE = hν − KE − φ. The software does this conversion for you, so spectra are plotted against binding energy, with the axis running from high BE on the left to low BE on the right.

Why is XPS surface sensitive?

X-rays travel micrometres into the sample, but the photoelectrons lose energy very quickly when they hit other atoms. Only electrons from the top few nanometres (about 3 inelastic mean free paths, typically < 10 nm) leave without losing energy and form the sharp peaks. Electrons that lost some energy on their way out land at higher apparent BE and build the rising background on the high-BE side of each peak.

top ≈ 10 nmX-rays (µm deep)escapes: sharp peakloses energy on the way:background, or never leaves
X-rays go micrometres deep, but only electrons from the top few nanometres escape without losing energy.

Why vacuum?

The photoelectrons must travel to the analyser without hitting gas molecules, and the surface must stay clean during the measurement. XPS therefore works in ultra-high vacuum (around 10⁻⁹ mbar).